Scope
& application
Overview
definitions
MR safety
Reasons
for initiating Quench
Health and
Safety Issues
Personnel
Qualifications
Location
of Systems and Switches
Quench Procedure
1.0 Scope & Application
1.1 This standard operating procedure is for an operator-initiated
quench on either the Trio or Sonata MR System. These
scanners are housed in the Basement of Rhoads and
Founders buildings at the Hospital of the University
of Pennsylvania. This SOP will cover:
1.1.1 A short review of safety in MR Environment
1.1.2 Reasons for initiating quench
1.1.3 Location of systems and switches
1.1.4 Actual Quench procedure
1.1.5 Potential issues during quench
1.1.6 Post Quench follow-up
2.0
Overview
2.1 Protecting the health and safety of PENN patients,
employees, and students is a top priority. This is
why working safely within the MR Research environment
is critical.
2.2 The success of this effort relies on the accountability
of each individual for their own and others health
and safety. This SOP is intended to provide direction
and information about an operator-initiated quench.
This procedure will rapidly (less than 1 minute) dissipate
the existing static field of the MR Magnet.
3.0
Definitions
3.1 Magnet - Super-conducting source of the static
field. This field strength 1.5 Tesla for the Sonata
and 3 Tesla for the Trio.
3.2
Static Field - This is the field generated by Super-conducting
magnet. This field is never turned off unless done
by field service or quenched.
3.3
Gauss -The centimeter-gram-second unit of magnetic
flux density, equal to one Maxwell per square centimeter.
10,000 Gauss are equal to 1 Tesla.
3.4
Gauss Lines - demarcation lines of specific field
strengths
3.5
Oxygen monitor- Sensor inside magnet room that monitors
the oxygen level.
3.6
Quench "Operator Initiated" - MR Operator
presses the quench switch to rapidly remove the static
magnetic field from the MR Suite.
3.7
Spontaneous Quench - The magnet loses the static field
without operator initiation. A spontaneous quench
can happen if:
3.7.1
Cryogen Levels are allowed to drop past a critical
level
3.7.2 A large piece of ferrous material becomes lodged
against the magnet
3.7.3 Manual removal of a large ferrous object from
the magnet
3.7.4 A cryogenic leak
4.0
MR Safety
4.1 The MR magnetic fields are always on. No ferromagnetic
material can be taken inside the scanner room. The
ability to know whether a material is ferromagnetic
through visual inspection is impossible. Therefore,
all equipment containing any type of metal must not
be taken into the magnet room. All equipment entering
the room must be screened and approved by the Technologist
or approved user.
4.2 Personnel must be screened by the MR Technologist
or approved user before entering the MRI room.
4.3 Electronic Equipment may not be taken into the
scanner room. Electronic equipment often has components
that are pulled toward a magnetic field. The safety
of the equipment must be determined by the MRI Technologist
or CAMRIS member before being taken in the scanner
room.
4.4 Safety inspection (6.3) of equipment is done prior
to the first procedure. Small ceramic permanent magnets
are available at each scanner to evaluate magnetic
properties of an object. Ask the Technologist for
access to this device.
4.5 Metallic carts are not permitted in the scanning
areas.
4.6 All metal must be removed from the animals prior
to procedures. (e.g. ID Tags, collars)
5.0
Reasons for Initiating Quench
5.1 Reasons for an operator to initiate the quench
sequence are discussed below. Examples are given with
each to better illustrate the concepts:
5.1.1
Catastrophic injury :
5.1.1.1
An oxygen tank is brought into the MR Suite. It becomes
airborne and pins a person against the magnet.
5.1.1.2 A patient on a wheelchair is brought into
the MR Suite. The chair becomes airborne pinning the
patient or other personnel against the magnet.
5.1.2
Potential for Catastrophic Injury :
5.1.2.1
A patient on a wheelchair is brought into the MR Suite.
The chair is pulled against the magnet while the patient
is still in the chair. There is no immediate injury
to anyone but the potential for injury is high.
5.1.2.2 An oxygen tank becomes airborne and lodges
against the magnet while a patient or personnel are
in the room. Again no immediate injury but the large
ferrous object is lodged against the magnet presenting
an high potential for injury.
5.1.2.3 A fire has erupted in the MR Suite and the
arrival of Fire Department and security personnel
is imminent.
Our
first concern is for the safety of patients and personnel
in the MR environment. Stability of ferrous objects
lodge on or near the magnet cannot be readily assessed.
We must assume that any slight change in their orientation
or disposition could have severe consequences. Therefore
we want operators of this system to be aware of the
quench option and to use it if necessary.
6.0 Health and Safety Issues
6.1 Risks associated with a quench are:
6.1.1 Asphyxia - if the gaseous boil off of the cryogens
does not vent properly room air is displaced and persons
may asphyxiate.
6.1.2 Cryogenic Burns - If during the boil off some
of the magnet cryogens spill out. Personal contact
with these substances could cause burns.
6.1.3 Crushing Injuries - as the magnetic field dissipates
any objects that were suspended by the magnet will
fall and could cause injury.
6.1.4 Increased pressure discomfort - during the quench
the pressure in the room may increase causing ear
discomfort
6.1.5 Noise Injury - quenches can be very loud and
injure hearing
7.0
Personnel Qualifications
7.1 All individuals operating within the MRI environment
will be trained Radiologic Technologists or approved
users and have reviewed procedural standards of safety.
All individuals operating the scanner will have user
approval status acquired through CAMRIS.
8.0
Location of Systems and Switches
8.1 The Hup5 magnet is located in the basement of
the Founders Building. The quench venting duct of
this system is on the roof directly above the small
conference room in the Radiologist MR Learning Center.
The venting system is checked and maintained by Siemens.
8.2 The quench switches and oxygen sensor for Hup5
are located in the MR suite as shown on the following
diagram.
8.3
The Hup6 magnet is located in the basement of the
Rhodes Building. The quench venting duct of this system
is located directly above the large conference room
in the Radiologist MR Learning Center. The venting
system is checked and maintained by Siemens.
8.4 The quench switches and oxygen sensor for Hup6
are located in the MR suite as shown in the diagram.
9.0
Quench Procedure
9.1 Open Scanner Room Door - the door of the scanner
room should be open during the quench to allow any
gases not vented to escape from the room. Improperly
vented gas in the MR Suite can cause increased pressure
on the door and make it impossible to open. It will
also displace breathable air from the room.
9.2
Evacuate personnel from scanner room - the MR Suite
and surrounding area should have patients and nonessential
personnel removed from the area before the quench
when practicable.
9.3 Announce intention to quench - the operator should
announce the intention to activate the quench procedure
to alert all individuals in the immediate area.
9.4
Lift Plastic strike cover on quench button - the quench
switch is covered by a protective plastic cover. It
must be lifted up away from the button before the
button can be activated.
9.5
Press and hold Quench button for a count of 3 - the
quench switch must be pressed firmly and held closed
for a short time to ensure the desired quench takes
place.
9.6
Monitor oxygen level detector for warning - there
is an oxygen sensor in each MR suite mounted at head
level. It activates the oxygen alarm outside of the
room when oxygen level has reached a level which cannot
sustain life.
9.7
Evaluate presence of Magnetic field after Quench -
using the gauss meters available at the entrance to
each scanner walk into the room and look for evidence
of the displacement of the meter's indicator. If the
indicator does not move the field is no longer active.
9.8
Resolve issues leading to quench - It is now possible
to enter the room with equipment and emergency personnel
to retrieve objects or implement activities that are
necessary to resolve the issues leading to the quench.
9.9
Notify Siemens Medical and Systems Manager.
9.9.1.1 Hup5 and Hup6 - Norman Butler - 610-209-8438
9.9.1.2 Siemens Medical - 1-800-888-7436