TESTING
OF PULSE SEQUENCES
All
research pulse sequences that are to be used on human
subjects must be approved by the CAMRIS committee
prior to human use. Formal approval from the CAMRIS
committee or its designee is required before use of
a research sequence on any volunteer, including the
researchers themselves. Research pulse sequences include
those written at HUP or any other institution or corporation.
The only sequences that are exempted are those provided
by the manufacturer as product and are used within
the parameters and protocols for which they were developed.
Overriding the design of a product sequence using
"control variables" requires CAMRIS approval.
Modification of any aspect of an approved sequence
requires CAMRIS re-approval before human use is allowed.
Additionally, disabling of the RF power monitor is
prohibited.
We
anticipate that the approval process will only require
the documentation of steps that are already being
performed by researchers in the development of their
sequences, so this process should not represent a
burden. Our goal is to formalize and document these
procedures. The approval procedure requires the following
documents to be submitted to the CAMRIS committee:
1.A
brief description of the sequence, paying particular
attention to the RF pulse parameters and duty cycle.
Include in this description the method of the SAR
estimation. SAR may be calculated using the RF pulse
structure and code within the pulse sequence, or may
be calculated separately. Comparing the new sequence
to an existing approved sequence can be a helpful
approach. If the sequence is to be used on the 4 Tesla
magnet only, then SAR calculations are not necessary
since a maximum power estimate has already been performed
for the coil approval.
2.
A copy of the actual pulse sequence source code on
diskette.
3.
A timing diagram of the pulse sequence. This diagram
should be generated by the sequence development tools
on the workstations. This will document the expected
sequence output. Please provide the value of the control
value ‘xmtaddScan’ and make sure that the appropriate
gradient configuration file (usually 1.5T_SR120_config.cvs)
is used when performing the sequence simulation. If
multiple ‘boffsets’ are used within the sequence,
provide a timing diagram for each sub-sequence. Please
use sufficient resolution so as to discern areas of
interest (for example,
gradient rise times) and turn on axis
labeling. Multiple pages may be required to show sufficient
detail.
4.
A phantom image using the new sequence.
5.
The imaging protocol for human use. This may include
a range of values to allow flexibility.
When
naming the sequences, use a numeric extension that
increments on each revision (for example, fgre0, fgre1,
… fgreN) so that modifications and their subsequent
approval can be documented. Each revision will require
CAMRIS approval. Many revisions may qualify for expedited
approval as described below.
Documents
will be submitted for approval at the CAMRIS meeting
following their submission. Sequence changes that
do not significantly alter the dB/dt or RF may be
reviewed by Larry Dougherty or Hee Kwon Song in an
expedited fashion and then presented at the next CAMRIS
meeting.
|